Druck LiquidTC 255 Liquid Bath Temperature Calibrator

FEATURES

Temperature range from Ambient to 255C.
Accuracy from 0.1°C.
Stability 0.05°C.
Large bath for irregular and multiple devices.
Multi-purpose – liquid bath, black body source, dry block.
Interchangeable bath – simplifies fluid changes.
Light weight and robust for field use.
Leak-proof bath cover for transportation.

หมวดหมู่: ,

Druck LiquidTC 255 Liquid Bath Temperature Calibrator

The Druck LiquidTC 255 combines the portability of a dry block calibrator with the flexibility of a liquid immersion bath to enable the testing and calibration of virtually any shape and size of devices.

The calibrators are factory configured as liquid baths and are provided with a bath cover to hold up to 5 devices while reducing heat loss from the surface of the liquid medium.

For transportation a leak-proof sealing cover is also provided as standard. Optionally the temperature calibrators can be configured with additional capabilities including interchangeable liquid baths, a black body source and dry block interchangeable inserts.

Druck LiquidTC 255 Specifications
Temperature Range Ambient to 255°C
(Ambient to 490 °F option).
Heat Well Dia 60 mm/depth 170 mm (150 mm working depth).
Measuring Zone 0 to 40 mm from the bottom of the insert.
Stability 0.05°C.
Accuracy
Standard Liquid Bath – LI 0.2°C.
Dry Block Option – DB 0.4°C.
Infrared Black Body Option – IR 0.5°C.
Black Body Emissivity 0.9994.
Display
Display Range 0 to 255°C.
Measurement Resolution 0.01° from -9.99 to 99.99 otherwise 0.1°C.
Setting Resolution 0.1°C.
Standard Features
Functions Controller off, manual control, operating service hours, set-point memory, rate control °C/min, test profile.
Power Requirements
Supply Voltage 100 to 240 V AC 50/60 Hz.
Power Consumption 1000 VA nominal.
Dimensions/Weight
Width 150 mm.
Height 330 + 68 mm.
Depth 270 mm.
Weight 7.5 kg.
Environmental
Ambient Temperature 10°C to 50°C (50°F to 122°F)
Specifications valid at 18 °C to 28 °C (65 °F to 82 °F) ambient temperature. Minimum and maximum temperatures may not be achieved over the full ambient temperature range.
Ambient Humidity To 80% RH (non-condensing).
Storage Temperature From -10 to 60°C (14 to 140°F).
Recommended Operating Environment Indoor use. Not for use in potentially explosive atmospheres.
Compliance EN6326 (EMC), EN61010 Oelectrical safety), EU directive (RoHS, REACH, WEEE).
Druck LiquidTC 255 warranty One-year.
Scroll to Top